Vivid i N ultrasound. Not marketed in the US. Used in ultrasound imaging and analysis in Fetal/Obstetrics; Abdominal/... — FDA record Z-0758-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Nov 10, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Vivid i N ultrasound. Not marketed in the US. Used in ultrasound imaging and analysis in Fetal/Obstetrics; Abdominal/GYN; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculo-skeletal Conventional and Superficial; Urology (including prostate); Transesophageal; Transrectal; Transvaginal; and Intraoperative (abdominal, thoracic, and vascular).
Reason
Potential for batteries that have not been replaced at 2 years, as recommended in the Service Manual, they can fail or possibly emit smoke, or catch fire.
Classification
Class II
Recalling firm
GE Medical Systems, LLC
Quantity
113 units
Agency-reported status
Ongoing
Source record identifier
Z-0758-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityNo US Distribution. Model Number H45561AC, System ID: DZ1237US02, AO1015US01, EG1843US01, EG5882US01, EG1621US04, EG1621US03, 2538978U01, 083038042857318, 083038602007811, 083038602826113, 0835384095, IQ1165US05, IQ1376US01, KE1049US01, KW1010US04, KW1102US04, KW1001US35, KW1011US21, KW1011US22, KW1006US14, KW1098US06, KW1020US15, KT1061US01, KT1062US02, LB1359US04, 1553UL1026, 0834380243, 0834384085, 0834384095, MA1251US13, MA1272US10, OM1042US03, OM1007US11, OM1007US12, PK1331US08, PK1331US06, PA1097US01, SA1201US02, SA1014US07, SA1337US05, SA1060US08, SA1147US05, SA1149US03, SA1051US09, SA2220US02, SA2220US01, SA2072US17, SA1028US06, SA1089US07, SA1138US05, SA2158US07, SA1340US02, SA1223US15, SA1002US08, SA1036US06, SA1025US03, SA2214US03, SA1272US10, SA1024US04, SA1101US04, SA1034US20, SA1249US11, SA1034US23, SA1058US60, SA1058US47, SA1058US59, SA1332US03, SA1030US05, SA1282US05, ZA2573US01, ZA2422US01, ZA2328US01, ZA2339US01, ZA2420US01, ZA1024US07, ZA2348US01, ZA2358US01, ZA1030US10, ZA2421US01, ZA1025US16, ZA1025US25, ZA2378US01, ZA2510US01, TN1028US04, 01002ULT25, 06004ULT19, 07021ULT02, P34189UL17, 21002ULT18, 23002ULT06, P23003UL12, 32004ULT18, 34013ULT13, 34037ULT16, 34013ULT14, P35076UL03, 38006ULT08, 42002ULT12, 42002ULT10, P50005UL02, 55001ULT13, 60001ULT07, P61029UL05, AE1283US01, AE1353US01, AE1016US16, AE1016US14, AE1016US15, AE1011US18, AE1331US01, AE1015US16, YE1071US04, YE1071US03

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV and the countries of Afghanistan, Albania, Algeria, Angola, Argentina, Armenia, Australia, Austria, AZERBAIJAN, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Bulgaria, Burkina Faso, Cambodia, Cameroon, Canada, Chile, CHINA Colombia, CONGO, Costa Rica, Cote D'Ivoire, Croatia, Cyprus, CZECH REPUBLIC, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, Finland, France, Georgia, Germany, Ghana, Greece, Guatemala, Honduras, HONG KONG, Hungary, ICELAND, INDIA, Indonesia, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea (Republic of), Kosovo, Kuwait, KYRGYZSTAN, Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, LUXEMBOURG, Macedonia, MALAYSIA, Mali, Mauritania, Mauritius, Mexico, MICRONESIA, Moldova, Morocco, Mozambique, Namibia, Nepal, Netherlands, NEW ZEALAND, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, SYRIA, TAIWAN, TANZANIA, THAILAND, Trinidad and Tobago, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Vietnam, Yemen.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Dec 28, 2022
Recall initiation date
Nov 10, 2022
Center classification date
Dec 22, 2022

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.