Cardinal Health MEDI-VAC CRD Suction Canister Liner, 1000 mL Suction Canister Liner Kit with 5 mm 1.8 m Tube, REF 656... — FDA record Z-0755-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Dec 21, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Cardinal Health MEDI-VAC CRD Suction Canister Liner, 1000 mL Suction Canister Liner Kit with 5 mm 1.8 m Tube, REF 65651-516
Reason
The above-listed lots of Cardinal Health Medi-Vac CRD" Suction Canister Liners 1L may experience loss of suction on low or low intermittent suction settings, due to a misalignment of the liner with the outer hard canister.
Classification
Class II
Recalling firm
Cardinal Health 200, LLC
Quantity
203244 units
Agency-reported status
Ongoing
Source record identifier
Z-0755-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicability50630140152633 (case), Lot Numbers: J308-606, J308-624, J308-628 J308-642, J308-643, J309-608, J309-610, J309-622, J309-626, J309-633, J309-645, J310-602, J310-617, J310-645

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US, Canada, Australia and New Zealand

No US distribution state could be established from this record. Check the official notice.

Dates in the source

Report date
Jan 24, 2024
Recall initiation date
Dec 21, 2023
Center classification date
Jan 18, 2024