Recall archive / FDA device enforcement
Calibration Serum Level 3. CAL 3, REF CAL2351. Human Assayed Multi-Sera Level 2 (HN1530) Human Assayed Multi-Sera Lev... — FDA record Z-0751-2024
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Nov 13, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Calibration Serum Level 3. CAL 3, REF CAL2351. Human Assayed Multi-Sera Level 2 (HN1530) Human Assayed Multi-Sera Level 3 (HE 1532)
- Reason
- There have been transcription errors on the Instructions For Use (IFU) of the Calibration Serum Level 3 (CAL2351 lot 1260UE, 1262UE and 1315UE) for target values of Bilirubin and Triglycerides, on the IFU of Human Assayed Multi-Sera Level 2 (HN1530) for TBIC taget value and on IFU of Human Assayed Multi-Sera Level 3 (HE 1532) for Lipase target value.
- Classification
- Class II
- Recalling firm
- Randox Laboratories Ltd.
- Quantity
- N/A
- Agency-reported status
- Ongoing
- Source record identifier
- Z-0751-2024
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | UDI-DI: 05055273200966, lot1260UE, exp. 28 Nov 2024; lot 1262UE, exp. 28 Jan 2025; and lot 1315UE, exp. 28 May 2025 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
US Nationwide distribution.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Jan 24, 2024
- Recall initiation date
- Nov 13, 2023
- Center classification date
- Jan 18, 2024