HeartWare HVAD Pump Kit, REF 1104CA-CLIN — FDA record Z-0747-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Nov 29, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
HeartWare HVAD Pump Kit, REF 1104CA-CLIN
Reason
Over time, the driveline cover may become hardened, creating difficulty or an inability to slide the cover back to access the driveline to controller connector. Access to the driveline connector may be delayed, which may result in patient harm from a prolonged pump stop, if an urgent driveline connection or controller issue needs to be addressed. In the event a hardened driveline cover is observed, contact your Medtronic field representative for further assessment and/or servicing.
Classification
Class II
Recalling firm
Heartware, Inc.
Quantity
6 units
Agency-reported status
Ongoing
Source record identifier
Z-0747-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilitySerial Numbers: HW24101, HW24102, HW24103, HW24104, HW24755, HW24763

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution.

No US distribution state could be established from this record. Check the official notice.

Dates in the source

Report date
Dec 28, 2022
Recall initiation date
Nov 29, 2022
Center classification date
Dec 21, 2022