ProteusPLUS and ProteusONE- designed to produce and deliver a proton beam for the treatment of patients with localize... — FDA record Z-0747-2022

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Feb 15, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
ProteusPLUS and ProteusONE- designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation Proteus 235
Reason
When resuming from a beam pause, the Proton Therapy System does not verify whether the beam range has not been manually modified during the pause and is still the prescribed one. Therefore, if an operator manually modified the range during a pause, there is a risk that a portion of the treatment beam after the resume is delivered with an error in range
Classification
Class II
Recalling firm
Ion Beam Applications S.A.
Quantity
15 units US; 22 units OUS
Agency-reported status
Ongoing
Source record identifier
Z-0747-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityPAT.003 (KR), PAT.006 (US), PAT.107 (EU), PAT.108 (US), PAT109 (US), PAT.110 (US), PAT.111 (EU), PAT.112 (US), PAT.113 (US), PAT.114 (EU), PAT.115 (EU), SAT.116 (US), SAT.117 (EU), SAT.118 (RU), SAT.119 (US), SAT.120 (EU), SAT.122 (EU), SAT.123 (US), SAT.125 (IN), SAT.126 (US), SAT.127 (TW), SAT.132 (EU), SAT.133 (US), SAT.136 (IN), SAT.140 (US), SBF.101 (EU), SBF.103 (JP), SBF.104 (JP), SBF.105 (US), SBF.107 (FR), SBF.109 (UK), SBF.110 (UK), SBF.112 (EU), SBF.113 (US), SBF.115 (UK), SBF.117 (EU), SBF.128 (US). UDI: (01)05404013801138

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

IL, KS, LA, FL, MI, OK, PA, NJ, TN, TX, VA, WA Foreign: Belgium, England, Germany, France, Czech Republic, Italy, Poland, Russia, Spain, Sweden, South Korea, Taiwan, Netherlands,

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Mar 16, 2022
Recall initiation date
Feb 15, 2022
Center classification date
Mar 10, 2022

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records