Galaxy System (GAL-001). Intended to provide bronchoscopic visualization of and access to patient airways for diagnos... — FDA record Z-0742-2025
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Oct 28, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Galaxy System (GAL-001). Intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures.
- Reason
- Bronchoscopic visualization system power supplies have defective semiconductor components that may fail unexpectedly, which could result in an aborted procedure since it is a non recoverable failure. Manual removal of the bronchoscope when the system is not functioning can increase the chance of injury to the tissues of the lung as the scope is withdrawn including the possibility of pneumothorax.
- Classification
- Class II
- Recalling firm
- Noah Medical
- Quantity
- 18
- Agency-reported status
- Ongoing
- Source record identifier
- Z-0742-2025
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | UDI-DI: 00850048825048. Serial Numbers: CRB-1502, CRB-1503, CRB-1506, CRB-1508, CRB-1516, CRB-1517, CRB-1519, CRB-1520, CRB-1521, CRB-1522, CRB-1523, CRB-1524, CRB-1525, CRB-1527, CRB-1528, CRB-1529, CRB-1530, CRB-1532, |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
US: SC, OH, PA, TN, MO, IN, CA, WV, NY, MD, MA, NJ, ID, CT, NC
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
Dates in the source
- Report date
- Jan 1, 2025
- Recall initiation date
- Oct 28, 2024
- Center classification date
- Dec 20, 2024
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.