PosiFlush Prefilled Saline Syringe with General Pump Compatibility (10mL Syringe with 10mL Saline Fill), Catalog Numb... — FDA record Z-0736-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Nov 16, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
PosiFlush Prefilled Saline Syringe with General Pump Compatibility (10mL Syringe with 10mL Saline Fill), Catalog Number 306547
Reason
An incorrect stopper was used in nine lots of 10mL BD PosiFlush Normal Saline Flush Syringes with General Pump Compatibility. The use of BD PosiFlush Prefilled Flush Syringes with this stopper causes an increase in injection force, which has the potential to trigger alarms if used with a syringe pump.
Classification
Class II
Recalling firm
Becton Dickinson & Company
Quantity
5,085,120 units
Agency-reported status
Ongoing
Source record identifier
Z-0736-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 30382903065470; Lot Numbers: 3207548 3207555 3214826 3226386 3214828 3226388 3250302 3250303 3250304

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution including in the states of AL, AZ, CA, CO, FL, GA, HI, IA, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX,UT, VA, WA, WI, WV.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jan 24, 2024
Recall initiation date
Nov 16, 2023
Center classification date
Jan 17, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.