2008T and 2008TBluestar Hemodialysis Machines with CDX-Indicated for acute and chronic dialysis therapy in a healthca... — FDA record Z-0734-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jul 27, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
2008T and 2008TBluestar Hemodialysis Machines with CDX-Indicated for acute and chronic dialysis therapy in a healthcare facility Models: (1) 190713 2008T HEMODIALYSIS SYS., with CDX; (2) 190766 2008T HEMODIALYSIS SYSTEM W/BIBAG W/CDX ; (3) 191124 2008T HD SYS. CDX BLUESTAR; (4) 191126 2008T HD SYS. CDX W/BIBAG BLUESTAR
Reason
(1)Software version 2.74 upgrade and (2)hardware related cause that affects only CDX Machines, has been identified as the UI-MICs Board is sensitive to Electromagnetic Interference (EMI) which will cause the false alarm "Remove USB 2 Device".
Classification
Class II
Recalling firm
Fresenius Medical Care Holdings, Inc.
Quantity
2008T with CDX= 70,552 2008T Bluestar with CDX= 25,205
Agency-reported status
Ongoing
Source record identifier
Z-0734-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilitySoftware issue in SW versions 2.72 and earlier with CDX. UDI: (1) 00840861100897; (2) 00840861100910; (3) 00840861102082; (4) 00840861102099

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Dec 28, 2022
Recall initiation date
Jul 27, 2022
Center classification date
Dec 20, 2022

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records