Recall archive / FDA device enforcement
Ultrasound EA (Ergonomic Advantage) table version; V2 (Vascular 2 Section) Echocardiography -An Ultrasound examinatio... — FDA record Z-0730-2024
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Dec 5, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Ultrasound EA (Ergonomic Advantage) table version; V2 (Vascular 2 Section) Echocardiography -An Ultrasound examination table
- Reason
- Backrest platform stress fracture - the table is not operational in an inclined position
- Classification
- Class II
- Recalling firm
- Oakworks Inc
- Quantity
- 28 units
- Agency-reported status
- Ongoing
- Source record identifier
- Z-0730-2024
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | (UDI): 00817463025521 S/N: VCF805597 VCF814488 VCF814750 VCF817222 VCF817223 VCF817720 VCF817721 VCF823450 VCF825629 VCF825630 VCF833350 VCF813968 VCF817520 VCF826533 VCF822117 VCF824501 VCF824502 VCF824503 VCF824504 VCF815120 VCF814849 VCF835728 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
US Nationwide Distribution
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Jan 24, 2024
- Recall initiation date
- Dec 5, 2023
- Center classification date
- Jan 12, 2024
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.