O&M Halyard Procedure Mask with SO SOFT* Earloops, Yellow-Disposable surgical mask Product Code: 47117 — FDA record Z-0728-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Nov 8, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
O&M Halyard Procedure Mask with SO SOFT* Earloops, Yellow-Disposable surgical mask Product Code: 47117
Reason
Not meeting the bioburden testing requirements of EN 14683 required for masks sold in Europe. Face masks sold in the United States are not required to meet this standard. However, if the standard is printed on the packaging, compliance is required.
Classification
Class II
Recalling firm
O&M HALYARD, INC.
Quantity
211 cases
Agency-reported status
Ongoing
Source record identifier
Z-0728-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 20680651471172 Lot Number: AM4049121 Exp. Date: 2029-02-18

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Nationwide

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jan 1, 2025
Recall initiation date
Nov 8, 2024
Center classification date
Dec 20, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.