VADER Pedicle System Torque Wrench, Hudson Coupling, 12 Nm, REF 42-702 — FDA record Z-0728-2023

Reported terminated

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Nov 3, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
VADER Pedicle System Torque Wrench, Hudson Coupling, 12 Nm, REF 42-702
Reason
During internal testing (at manufacturer site) of two torque wrench devices, it was detected that the torque limiting function was out of specification. The specified torque limiting value is 12 Nm. The measuring results for both instruments were 4 Nm.
Classification
Class II
Recalling firm
Icotec Ag
Quantity
6 devices
Agency-reported status
Terminated
Source record identifier
Z-0728-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI/DI: (01) 07640164844589, Serial Numbers: 200221-040, 200221-041, 200221-042, 200221-043, 200221-044, 200221-045

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the state of GA.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Dec 28, 2022
Recall initiation date
Nov 3, 2022
Center classification date
Dec 19, 2022
Termination date
Sep 10, 2026