Incisive CT, Model Numbers 728144 and 728143. Computed Tomography X-Ray System PIM cable part number 459801179871 — FDA record Z-0719-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Dec 7, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Incisive CT, Model Numbers 728144 and 728143. Computed Tomography X-Ray System PIM cable part number 459801179871
Reason
If the Patient Interface Monitor (PIM) Cable is plugged/unplugged frequently as part of daily operation, the connector and wire may disconnect from the welding point, which may result in intermittent loss of the ECG signal. Loss of ECG signal may lead to the decision to rescan the patient or delay in diagnosis during a clinical emergency.
Classification
Class II
Recalling firm
Philips North America
Quantity
55 US; 1029 ROW
Agency-reported status
Ongoing
Source record identifier
Z-0719-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityIncisive CT systems configured with the PIM cable option are affected by this issue. To determine whether your system is affected, please locate the connection cable between your ECG monitor and incisive CT gantry and verify that the part number on your PIM cable matches 459801179871 . (1) REF 728143 - UDI-DI 00884838085015 Serial Numbers: 33007 33038 33013 33006 33012 33045 33011 500102 33015 33005 33041 33047 500217 500255 33035 500252 500145 500046 500168 33009 33046 33024 500222 500064 33034 500242 500301 33010 33019 33001 33020 33031 33037 500110 500111 500178 33003 500282 33002 33033 (2) REF 728144 - UDI-DI 00884838105508 Serial Numbers: 34002 34051 34050 34041 34059 34057 34054 34040 34007 34026 34058 34049 34056 34052

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Domestic distribution to the following states: AL CA CO FL GA HI IL IN KS KY LA MD MI MO MS ND NE NJ NY OH OK PA RI SC TX UT VT WA International distribution worldwide.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jan 24, 2024
Recall initiation date
Dec 7, 2023
Center classification date
Jan 12, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.