Everest MI XT Inner Dilator, Catalog Number 5101-90167 — FDA record Z-0705-2022

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jan 13, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Everest MI XT Inner Dilator, Catalog Number 5101-90167
Reason
Stryker received two (2) complaints for units from specific lots of Inner and Outer Dilators not fitting together properly during a surgical procedure. Some dilators from these lots and from other lots were subsequently determined to be affected by a manufacturing nonconformance.
Classification
Class II
Recalling firm
K2M, Inc
Quantity
686 Total Devices
Agency-reported status
Ongoing
Source record identifier
Z-0705-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityGTIN: 10888857261204; Lot Numbers: JUJF KFMV KUPG KYYU MDRB NDMK NDMX NXJK

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Mar 9, 2022
Recall initiation date
Jan 13, 2022
Center classification date
Mar 1, 2022

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.