Brand Name: K-Systems Product Name: Multiple/unknown Model/Catalog Number: Multiple/unknown Software Version: N/A Pro... — FDA record Z-0692-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Nov 15, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Brand Name: K-Systems Product Name: Multiple/unknown Model/Catalog Number: Multiple/unknown Software Version: N/A Product Description: An electrically-powered device intended to provide suitable warming and/or incubation conditions for reproductive biological materials (e.g., gametes and embryos), at closely regulated and monitored temperatures, during assisted reproduction examination and processing procedures. It may support a more comprehensive growth environment [e.g., temperature, carbon dioxide (CO2) and oxygen (O2) concentration]; it typically has adjustment and monitoring controls. Component: N/A
Reason
The warming surfaces or incubator well temperatures of the listed products may exceed the desired set temperature. If temperatures exceed the desired set value, the device will alarm appropriately.
Classification
Class II
Recalling firm
CooperSurgical, Inc.
Quantity
34 units
Agency-reported status
Ongoing
Source record identifier
Z-0692-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUnknown product distributed in India and Australia as lab equipment. Version or Model: Multiple/unknown; Primary DI Number: unknown; Serial Numbers: unknown. NOTE: These devices are one of the five (5) noted affected Part Numbers, so the intended use, etc. will be one of those other devices. As stated above, traceability was not a concern due to the classification of these devices in those regions.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

States: NC, NJ, TX, VA, IL, NY, CA, PA Please note: The countries below, excluding Iraq, all consider these devices to be "lab equipment", not medical devices and are not subject to classification or recall guidelines. Countries: AT, AU, BE, BH, BR, CA, CH, CN, CZ, DE, EG, ES, ET, FR, GB, GH, HU, ID, IE, IL, IN, IQ, IT, JO, JP, KR, KW, LB, MA, MK, MT, NL, NO, PE, PH, PK, PT, QA, RO, RS, SA, SD, SE, SO,TR, TW, TZ, US, ZA

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Dec 25, 2024
Recall initiation date
Nov 15, 2024
Center classification date
Dec 18, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.