Recall archive / FDA device enforcement
Brand Name: K-Systems Product Name: T47 Warming Plate Model/Catalog Number: K23054 Software Version: N/A Product Desc... — FDA record Z-0689-2025
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Nov 15, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Brand Name: K-Systems Product Name: T47 Warming Plate Model/Catalog Number: K23054 Software Version: N/A Product Description: An electrically-powered device intended to provide suitable warming and/or incubation conditions for reproductive biological materials (e.g., gametes and embryos), at closely regulated and monitored temperatures, during assisted reproduction examination and processing procedures. It may support a more comprehensive growth environment [e.g., temperature, carbon dioxide (CO2) and oxygen (O2) concentration]; it typically has adjustment and monitoring controls. Component: No
- Reason
- The warming surfaces or incubator well temperatures of the listed products may exceed the desired set temperature. If temperatures exceed the desired set value, the device will alarm appropriately.
- Classification
- Class II
- Recalling firm
- CooperSurgical, Inc.
- Quantity
- 22 units
- Agency-reported status
- Ongoing
- Source record identifier
- Z-0689-2025
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | Version or Model: K23054; Primary DI Number: 00888937025071; Serial Numbers: 2218KH46 , 2218KH47 , 2218KH48 , 2218KH49 , 2218KH50 , 2218KH51 , 2218KH52 , 2218KH53 , 2218KH54 , 2218KH55 , 2218KH56 , 2218KH57 , 2218KH58 , 2218KH59 , 2218KH60 , 2233KH11 , 2233KH12 , 2233KH13 , 2233KH14 , 2233KH15 , 2233KH16 , 2233KH17 , 2233KH18 , 2233KH19 , 2233KH20 , 2235KH31 , 2235KH32 , 2235KH33 , 2235KH34 , 2235KH35 , 2235KH36 , 2235KH37 , 2235KH38 , 2235KH39 , 2235KH40 , 2237KH11 , 2237KH12 , 2237KH13 , 2237KH14 , 2237KH15 , 2237KH16 , 2237KH17 , 2237KH18 , 2237KH19 , 2237KH20 , 2242KH01 , 2242KH02 , 2242KH03 , 2242KH04 , 2242KH05 , 2242KH06 , 2242KH07 , 2242KH08 , 2242KH09 , 2242KH10 , 2250KH01 , 2250KH02 , 2250KH03 , 2250KH04 , 2250KH05 , 2250KH06 , 2250KH07 , 2250KH08 , 2250KH09 , 2250KH10 , 2309KH41 , 2309KH42 , 2309KH43 , 2309KH44 , 2309KH45 , 2309KH46 , 2309KH47 , 2309KH48 , 2309KH49 , 2309KH50 , 2320KH01 , 2320KH02 , 2320KH03 , 2320KH04 , 2320KH05 , 2320KH06 , 2320KH07 , 2320KH08 , 2320KH09 , 2320KH10 , 2322KH01 , 2322KH02 , 2322KH03 , 2322KH04 , 2322KH05 , 2322KH06 , 2322KH07 , 2322KH08 , 2322KH09 , 2322KH10 , 2337KH01, 2337KH02, 2337KH03, 2337KH04, 2337KH05, 2337KH06, 2337KH07, 2337KH08, 2337KH09, 2337KH10, |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
States: NC, NJ, TX, VA, IL, NY, CA, PA Please note: The countries below, excluding Iraq, all consider these devices to be "lab equipment", not medical devices and are not subject to classification or recall guidelines. Countries: AT, AU, BE, BH, BR, CA, CH, CN, CZ, DE, EG, ES, ET, FR, GB, GH, HU, ID, IE, IL, IN, IQ, IT, JO, JP, KR, KW, LB, MA, MK, MT, NL, NO, PE, PH, PK, PT, QA, RO, RS, SA, SD, SE, SO,TR, TW, TZ, US, ZA
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
Dates in the source
- Report date
- Dec 25, 2024
- Recall initiation date
- Nov 15, 2024
- Center classification date
- Dec 18, 2024
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.