StatStrip Glucose Hospital Meter System. Model/Catalog Number: 63685. Prescription use blood glucose meter for near-p... — FDA record Z-0681-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Nov 20, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
StatStrip Glucose Hospital Meter System. Model/Catalog Number: 63685. Prescription use blood glucose meter for near-patient testing.
Reason
A software bug within the Gen 2 StatStrip Hospital Meter firmware leads to transmission of erroneous glucose patient test results to a healthcare institution's data management system (DMS) at healthcare sites using WiFi if a user navigates to the Review Results screen while the meter is in the process of transmitting the current glucose test result to the DMS. The risk of this software defect is erroneously high or low glucose results being documented in a patient's medical record, which may lead to incorrect treatment.
Classification
Class I
Recalling firm
Nova Biomedical Corporation
Quantity
2,365 units
Agency-reported status
Ongoing
Source record identifier
Z-0681-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityModel No 63685. UDI-DI (01)10385480636858. Software Versions: v0.0.13.10 to v0.0.13.44

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Distribution US Nationwide including Puerto Rico, Germany, Italy, Israel, Switzerland, Australia, Great Britain, and the Netherlands.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jan 1, 2025
Recall initiation date
Nov 20, 2024
Center classification date
Dec 20, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records