Comprehensive Shoulder System Primary Shoulder Stem, Micro Length, Porous Plasma, 11 MM, 55 MM Long Item Number: 113611 — FDA record Z-0671-2022

Reported terminated

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jan 7, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Comprehensive Shoulder System Primary Shoulder Stem, Micro Length, Porous Plasma, 11 MM, 55 MM Long Item Number: 113611
Reason
The products in scope underwent an incorrect rework operation for the porous plasma sprayed coating. Internal testing indicated that the parts may not have sufficient adhesion strength. The scope was identified to be all lots that were porous plasma coated or recoated at a vendor that were then reworked through the internal waterjet rework process. The waterjet rework process for porous plasma coating is not validated for products that were coated at a vendor.
Classification
Class II
Recalling firm
Biomet, Inc.
Quantity
10 units
Agency-reported status
Terminated
Source record identifier
Z-0671-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityLot Number: 821170 UDI: (01)00880304554122(17) 280825(10)821170

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide and the countries of Canada, Ecuador, Netherlands.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Mar 2, 2022
Recall initiation date
Jan 7, 2022
Center classification date
Feb 24, 2022
Termination date
Apr 13, 2023