MICRO-X Rover Mobile X-ray System, # MXU-RV19 — FDA record Z-0666-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Oct 23, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
MICRO-X Rover Mobile X-ray System, # MXU-RV19
Reason
During internal testing it was found that if the system is Ready to expose and the hand switch is then dropped into the stowage unit such that it lands on the prep/expose switch, the prep switch can be activated for a very short time such that the orange ring of death is displayed in the tube cooling area of the screen. The system then becomes unresponsive to kV/mAs changes, prep/expose presses. The only way to proceed is to suspend the patient, Exit Nexus DR and relogin and continue the exam.
Classification
Class II
Recalling firm
Micro-X Ltd.
Quantity
23
Agency-reported status
Ongoing
Source record identifier
Z-0666-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 09357123000013. Serial Numbers: 262, 274, 320, 329, 237, 269, 271, 290, 292, 307, 310, 311, 312, 321, 322, 323, 327, 330, 331, 332, 333, 234, 266.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide Distribution: TX, CA, OH, ID, WA, AZ

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Dec 18, 2024
Recall initiation date
Oct 23, 2024
Center classification date
Dec 12, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.