Recall archive / FDA device enforcement
MICRO-X Rover Mobile X-ray System, # MXU-RV35 — FDA record Z-0664-2025
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date May 31, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- MICRO-X Rover Mobile X-ray System, # MXU-RV35
- Reason
- Shots were terminated by the mAs integrator; however, the 4 ms exposure time did not allow enough tolerance to achieve diagnostic exposures due to variations in hardware and how different timers are started.
- Classification
- Class II
- Recalling firm
- Micro-X Ltd.
- Quantity
- 5
- Agency-reported status
- Ongoing
- Source record identifier
- Z-0664-2025
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | UDI-DI: 9357123000037. Serial Numbers: 00334, 00348, 00349, 00351, 00361.¿ |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Worldwide and US Nationwide Distribution: NJ, PR, CA
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Dec 18, 2024
- Recall initiation date
- May 31, 2023
- Center classification date
- Dec 12, 2024