MICRO-X Rover Mobile X-ray System, # MXU-RV35 — FDA record Z-0664-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date May 31, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
MICRO-X Rover Mobile X-ray System, # MXU-RV35
Reason
Shots were terminated by the mAs integrator; however, the 4 ms exposure time did not allow enough tolerance to achieve diagnostic exposures due to variations in hardware and how different timers are started.
Classification
Class II
Recalling firm
Micro-X Ltd.
Quantity
5
Agency-reported status
Ongoing
Source record identifier
Z-0664-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 9357123000037. Serial Numbers: 00334, 00348, 00349, 00351, 00361.¿

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide and US Nationwide Distribution: NJ, PR, CA

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Dec 18, 2024
Recall initiation date
May 31, 2023
Center classification date
Dec 12, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.