The Optima Coil System consists of an implantable embolization coil comprised of a platinum-tungsten alloy attached t... — FDA record Z-0660-2022

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jan 4, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
The Optima Coil System consists of an implantable embolization coil comprised of a platinum-tungsten alloy attached to a proximal stainless steel hypo-tube and distal body coil delivery pusher with a radiopaque distal positioning marker and proximal fluorosafe marker. The Optima Coil delivery pusher is 185cm in length. it is designed for use with the XCEL Detachment Controller. The XCEL Detachment Controller comes pre-loaded with batteries and is a sterile, handheld, single-patient-use device which should not to be reused, re-sterilized, opened or tampered with. The Optima Coil System and XCEL Detachment Controller(s) are sold separately. The Optima Coil achieves detachment by an internal heater element, which is powered by the XCEL Detachment Controller.
Reason
Due to Label/Pouch mix-up. Pouches incorrectly labeled. Incorrect shelf carton label or pouch label due to inadequate manufacturing line clearance. As a result of this labeling mix-up, there is the possibility of inadvertently selecting the incorrect coil size for the procedure potentially causing some harms like vessel damage, an inability to treat with said device and potential rupture of the aneurysm.
Classification
Class II
Recalling firm
BALT USA, LLC
Quantity
51 devices
Agency-reported status
Ongoing
Source record identifier
Z-0660-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityModel/REF: OPTI0204CSF10 Lot Number: F210500134 UDI Code: 00818053026010 Model/REF: OPTI0407CSF10 Lot Number: F210500135 UDI Code: 00818053026096

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - U.S. Nationwide distribution in the states of FL and MD. The countries of Denmark, France, Germany, Lebanon, Netherlands, Sweden, Peru, Portugal, Slovenia, Spain, and United Arab Emirates.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Mar 2, 2022
Recall initiation date
Jan 4, 2022
Center classification date
Feb 22, 2022