Vitality Polyaxial Screw, 7.5mmx 50mm, REF: 07.02000.118, used with Vitality and Vital Spinal Fixation System — FDA record Z-0639-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Oct 16, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Vitality Polyaxial Screw, 7.5mmx 50mm, REF: 07.02000.118, used with Vitality and Vital Spinal Fixation System
Reason
Spinal fixation system 7.5mmx 50mm screws are packaged with the correct labeling, but the screws themselves were laser marked with an incorrect size description of 7.5 x 40; use of incorrect screws could lead to screw perforation and indention of vascular or other structures.
Classification
Class II
Recalling firm
Zimmer Biomet Spine Inc.
Quantity
3
Agency-reported status
Ongoing
Source record identifier
Z-0639-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 00889024002159, Lot: W834061

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the states of MN, NE.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jan 10, 2024
Recall initiation date
Oct 16, 2023
Center classification date
Jan 4, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.