Trilogy Evo and Trilogy Evo O2 Material Numbers BL2110X15B BR2110X18B CA2110X12B DE2100X13B DE2110X13B DS2100X11B DS2... — FDA record Z-0638-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Nov 27, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Trilogy Evo and Trilogy Evo O2 Material Numbers BL2110X15B BR2110X18B CA2110X12B DE2100X13B DE2110X13B DS2100X11B DS2110X11B EE2100X15B EE2110X15B ES2100X15B ES2110X15B EU2100X15B EU2100X19 EU2110X15B FR2100X14B FR2110X14B FX2100X15B GB2110X15B IA2100X15B IA2110X15B IN2100X15B IN2100X19 IN2110X15B IT2100X21B IT2110X21B JP2100X16B JP2110X16B KR2110X15B LA2100X15B LA2110X15B ND2110X15B RDE2100X13B RDE2110X13B RDS2110X11B RIN2100X15B RIN2110X15B RJP2100X16B RJP2110X16B UDS2110X11B
Reason
The following was missing from the Contraindications Statement: The AVAPS-AE therapy mode is contraindicated for invasive use and patients less than 10 kg. The AVAPS feature is contraindicated for patients less than 10 kg. Potential harms if devices are used on contraindicated patients include barotrauma, hypoventilation/hypercapnia, and rebreathing of excessive CO2.
Classification
Class II
Recalling firm
Philips Respironics, Inc.
Quantity
100,126 (54,691 US; 45,435 OUS) in total
Agency-reported status
Ongoing
Source record identifier
Z-0638-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityMaterial UDI-DI/GTIN BL2110X15B 606959055520 BR2110X18B 606959055476 CA2110X12B 606959052307 DE2100X13B 606959054103 DE2110X13B 606959054110 DS2100X11B 606959051997 DS2110X11B 606959051942 EE2100X15B 606959055698 EE2110X15B 606959055544 ES2100X15B 606959054097 ES2110X15B 606959054080 EU2100X15B 606959055674 EU2100X19 606959055681 EU2110X15B 606959054981 FR2100X14B 606959055643 FR2110X14B 606959055506 FX2100X15B 606959061019 GB2110X15B 606959054127 IA2100X15B 606959055636 IA2110X15B 606959055490 IN2100X15B 606959054059 IN2100X19 606959055575 IN2110X15B 606959051959 IT2100X21B 606959055438 IT2110X21B 606959055421 JP2100X16B 606959055582 JP2110X16B 606959051973 KR2110X15B 606959055483 LA2100X15B 606959055599 LA2110X15B 606959055452 ND2110X15B 606959055513 RDE2100X13B RDE2110X13B RDS2110X11B 606959060517 RIN2100X15B RIN2110X15B RJP2100X16B RJP2110X16B UDS2110X11B 606959060500 All units of listed models are affected by this correction.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution: US (nationwide)-AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, Including PR; and OUS (International) including countries of: AE, AR,AT, AU, BD, BE, BG, BO, BR, CA, CH, CL, CN, CO, CZ, DE, DK, DZ, ;EC EE, EG, ES, FI, FR, GB, GE, GF, GR, GT, HK, HR, HU, IE, IL, IN, IT, JO, JP, KE, KR, KW, LB, LT, MA, ME, MK, MM, MQ, MT, MX, MY, NG, NL, NO, NP, OM, PA, PE, PF, PH, PK, PL, PR, PT, PY, QA, RE, RO, RS, RU, SA, SE, SG, SI, SK, TH, TN, TR, TW, UA, YT, ZA and ZW.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jan 10, 2024
Recall initiation date
Nov 27, 2023
Center classification date
Jan 4, 2024