Recall archive / FDA device enforcement
Universal Chuck-The Universal Chuck is an Instrument Handle provided as an alternative instrument available for use i... — FDA record Z-0634-2022
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Dec 22, 2021. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Universal Chuck-The Universal Chuck is an Instrument Handle provided as an alternative instrument available for use in the TN-ADVANCED" Tibial Nailing System (TNA), TFN-ADVANCED" Proximal Femoral Nailing System (TFNA), RFN-ADVANCED" Retrograde Femoral Nailing System (RFNA), FRN-ADVANCED" Femoral Recon Nailing System (FRNA), and Flexible Monobloc Reamers. Instrument Handles are intended for holding instruments for manipulation during orthopedic surgery. Product No.: 03.043.001
- Reason
- Top cap may loosen and detach if the Universal Chuck becomes jammed and the user attempts to manually free the device, internal ball bearings may fall out onto the surgical field and cause surgical delay or, and if not recovered, adverse tissue reaction and infection.
- Classification
- Class II
- Recalling firm
- Synthes (USA) Products LLC
- Quantity
- 600 units
- Agency-reported status
- Ongoing
- Source record identifier
- Z-0634-2022
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | Lot/Serial No.: 20013101, 20052701, 20228901, 20229301, 20253901, 20558602, 20698102, 20777201 21869701, 21983904, 21995301 GTIN: 10886982297372 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
US Nationwide distribution.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Feb 16, 2022
- Recall initiation date
- Dec 22, 2021
- Center classification date
- Feb 10, 2022
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.