ATTUNE REVISION POSTERIOR FEMORAL AUGMENT SZ 9 12MM-intended as a Component in total knee arthroplasty (TKA) revision... — FDA record Z-0631-2022

Reported terminated

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jan 5, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
ATTUNE REVISION POSTERIOR FEMORAL AUGMENT SZ 9 12MM-intended as a Component in total knee arthroplasty (TKA) revision surgeries Product Number: 154909003
Reason
May have adhesive residue on the posterior surface of the femoral augment, if undetected prior to implantation, soft tissue irritation and/or infection may occur.
Classification
Class II
Recalling firm
DePuy Orthopaedics, Inc.
Quantity
6 units
Agency-reported status
Terminated
Source record identifier
Z-0631-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityLots: JH9651, JH9649 GTIN: 10603295383178

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the states of TN, TX.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Feb 16, 2022
Recall initiation date
Jan 5, 2022
Center classification date
Feb 9, 2022
Termination date
Dec 22, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.