SEROLOGY ToRCH lgM 2022 POSITIVE CONTROLI Intended for use with in vitro assays for determination of IgM antibodies t... — FDA record Z-0624-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Nov 6, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
SEROLOGY ToRCH lgM 2022 POSITIVE CONTROLI Intended for use with in vitro assays for determination of IgM antibodies to Cytomegalovirus (CMV), IgM antibodies to Rubella Virus, IgM antibodies to Toxoplasma gondii, IgM antibodies to Herpes Simplex Virus Type 1 and 2. Catalog Number: SR10349
Reason
A decrease in the reactivity of HSV Type 1/2 lgM in the Serology ToRCH lgM Positive Control when tested on the DiaSorin Liaison XL.
Classification
Class III
Recalling firm
Randox Laboratories Ltd.
Quantity
1 unit
Agency-reported status
Ongoing
Source record identifier
Z-0624-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityGTIN: 05055273216424 Bath/Lot Number: 157SR

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

NY, WV

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jan 10, 2024
Recall initiation date
Nov 6, 2023
Center classification date
Jan 2, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.