bellavista 1000 ventilator, Catalog #301.100.030 and bellavista 1000e ventilator, 301.100.130. The bellavista 1000/10... — FDA record Z-0622-2022

Reported terminated

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Dec 20, 2021. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
bellavista 1000 ventilator, Catalog #301.100.030 and bellavista 1000e ventilator, 301.100.130. The bellavista 1000/1000e ventilator is intended to provide positive pressure ventilatory support to adult and pediatric patients, and optionally, infant and neonatal patients.
Reason
Potential cessation in ventilation can occur under specific conditions.
Classification
Class I
Recalling firm
Vyaire Medical
Quantity
4189 units
Agency-reported status
Terminated
Source record identifier
Z-0622-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityHardware generation G6, software version V6.0.1600.0 or higher installed, all serial numbers. UDI 07640149381115 (bellavista 1000 ventilator) and 07640149388879 (bellavista 1000e ventilator).

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide distribution in the states of AL, CA, CO, CT, DC, FL, GA, ID, IL, IN, LA, MA, MD, MI, MO, MS, NC, NJ, NY, OH, OK, PA, SD, TX, VA, WA, WI and WV, including Puerto Rico and the countries of Argentina, Australia, Austria, Bahrain, Belgium, Bosnia and Herzegovina, Brazil, Brunei, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Dominican Republic, Egypt, El Salvador, France, Germany, Guatemala, India, Indonesia, Iran, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Latvia, Lithuania, Malaysia, Mexico, Namibia, Netherlands, Nicaragua, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, South Africa, South Korea, Spain, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom, Vietnam, and Zimbabwe.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Feb 23, 2022
Recall initiation date
Dec 20, 2021
Center classification date
Feb 16, 2022
Termination date
Aug 28, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.