Landauer microSTARii reader, Medical Dosimetry System, Model Numbers: 18000-000/18001-000/18007-000 — FDA record Z-0616-2024

Reported completed

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Nov 15, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Landauer microSTARii reader, Medical Dosimetry System, Model Numbers: 18000-000/18001-000/18007-000
Reason
MicroSTARii readers may be contributing to the nanodot dosimeter measurement inaccuracy in two possible ways. The first is related to the interaction of the reader s LED beam profile with the nanoDot dosimeter. The second is debris formation due to drawer actuation, resulting in a change of reader response that may not be detectable unless known dose QCs are being run regularly.
Classification
Class II
Recalling firm
Landauer
Quantity
584 units
Agency-reported status
Completed
Source record identifier
Z-0616-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityAll serial numbers.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US and Australia, Brazil, Canada, China, France, Germany, India, Japan, Mexico, Rep of Korea, Taiwan, Thailand, Viet Nam

No US distribution state could be established from this record. Check the official notice.

Dates in the source

Report date
Jan 3, 2024
Recall initiation date
Nov 15, 2023
Center classification date
Dec 22, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.