ZEUS ELISA Parvovirus B19 IgM Test System, Product Number SM9Z7701G — FDA record Z-0603-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Oct 31, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
ZEUS ELISA Parvovirus B19 IgM Test System, Product Number SM9Z7701G
Reason
The organization completed a product improvement process to the conjugate found within the certain ELISA kits. This modification was not reviewed and approved by FDA prior to its implementation.
Classification
Class II
Recalling firm
Zeus Scientific, Inc.
Quantity
293 kits
Agency-reported status
Ongoing
Source record identifier
Z-0603-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI: (01)00845533001724(17)240731(10)23030069 (01)00845533001724(17)241231(10)23060238 (01)00845533001748(17)241231(10)23060239; Lot Numbers: 23030069 23060238 23060239

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US States: CA, MA, MI, NJ, NY, TN, UT.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Dec 27, 2023
Recall initiation date
Oct 31, 2023
Center classification date
Dec 21, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records