Essenz HLM (Heart-Lung Machine), REF: 49-00-10, with Software — FDA record Z-0571-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Oct 18, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Essenz HLM (Heart-Lung Machine), REF: 49-00-10, with Software
Reason
Heart-Lung Machine GUI touchscreen may go dark for 10-15 seconds, safety/performance functions remain operational before reset, control maintained through backup control panel, but second reset may occur, may lead to gas blender standby, then gas blender must be reactivated using user interface on blender, sensors deactivated until new case started; GUI reset can lead to Hypoxia and Hypoperfusion.
Classification
Class II
Recalling firm
LivaNova Deutschland GmbH
Quantity
38
Agency-reported status
Ongoing
Source record identifier
Z-0571-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 04033817903062. Software Version: 1.5. Serial Numbers: 49PK00436, 49PK00108, 49PK00374, 49PK00479, 49PK00480, 49PK00481, 49PK00093, 49PK00091, 49PK00092, 49PK00163, 49PK00497, 49PK00498, 49PK00219, 49PK00223, 49PK00224, 49PK00225, 49PK00226, 49PK00241, 49PK00255, 49PK00292, 49PK00291, 49PK00293, 49PK00490, 49PK00487, 49PK00488, 49PK00278, 49PK00274, 49PK00277, 49PK00275, 49PK00276, 49PK00273, 49PK00457, 49PK00458, 49PK00459, 49PK00455, 49PK00456, 49PK00290, 49PK00288

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the states of CA, VA, GA, IN, NJ, FL, CO, MS, NV.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Dec 4, 2024
Recall initiation date
Oct 18, 2024
Center classification date
Nov 27, 2024