ARIES HSV 1&2 Assay, Catalog/REF# 50-10017 — FDA record Z-0553-2022

Reported terminated

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Nov 17, 2021. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
ARIES HSV 1&2 Assay, Catalog/REF# 50-10017
Reason
Impacted lot may leak inside the ARIES instrument. Testing confirmed a leak rate of 0.9%. Leaking cassettes could potentially lead to incorrect results or a delay in reporting results due to an Invalid.
Classification
Class II
Recalling firm
Luminex Corporation
Quantity
329
Agency-reported status
Terminated
Source record identifier
Z-0553-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityLot Numbers AB3600A and AB3560A

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide distribution in the states of New Jersey, California, Connecticut, Minnesota, Missouri, New Jersey, New Mexico, Pennsylvania, Tennessee, Washington, and Wisconsin) and the country of Malaysia.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Feb 9, 2022
Recall initiation date
Nov 17, 2021
Center classification date
Feb 2, 2022
Termination date
Jul 8, 2024