OXY-1 System, Model Number 001-0500-001. Used during cardiopulmonary bypass. OXY-1 System consists of Console, Pump D... — FDA record Z-0550-2022

Reported terminated

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Dec 14, 2021. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
OXY-1 System, Model Number 001-0500-001. Used during cardiopulmonary bypass. OXY-1 System consists of Console, Pump Drive, E Drive, and Pole Mount. The disposable Pump Lung Unit PLU part number 001-0400-100 is sold separately and is not part of this recall.
Reason
The firm has received complaints of the OXY-1 System Console experiencing a power interruption while in use, which disrupts therapy delivered to a patient on support. Disruption of therapy could lead to prolonged hypoxia.
Classification
Class I
Recalling firm
Abiomed, Inc.
Quantity
42 Units
Agency-reported status
Terminated
Source record identifier
Z-0550-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI 00860001797909 Software Version Number V-1.3.15.0, V-1.4.6.0 and V.1.4.8.0 Serial Numbers 20200231-04 20200249-01 20200249-05 20200309-02 20200339-01 20200339-03 20200352-01 20200442-01 20200442-02 20200442-05 20210768-01 20210795-02 20210870-01 20210880-01 20210890-01 20210894-01 20210895-01 20210896-01 20210897-01 20210795-04

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US distribution to FL, GA, IN, MD, NC, PA, TX.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Feb 9, 2022
Recall initiation date
Dec 14, 2021
Center classification date
Feb 3, 2022
Termination date
Feb 28, 2023