VITROS 4600 Chemistry System Product Code 6802445 and Refurbished Product Code 6900440 Running Software Version 3.8.0 — FDA record Z-0544-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Oct 10, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
VITROS 4600 Chemistry System Product Code 6802445 and Refurbished Product Code 6900440 Running Software Version 3.8.0
Reason
During planned monitoring of data post-release of Software Version 3.8.0 an increased occurrence of software issues relating to unresponsive subsystems, unresponsive user interface (sometimes referred to as a screen freeze), and/or printer disconnections was identified. The consequence of these issues would be a delay in reporting results due to the required system shutdowns to recover the system. This delay in reporting results could occur during any assay, including which the speed of the result is deemed critical or STAT and could result in a delay of treatment in a patient. Due to the release strategy of this software version, the user could not update the system but a field engineer had to manually install each affected system.
Classification
Class II
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Quantity
51 units
Agency-reported status
Ongoing
Source record identifier
Z-0544-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityProduct Code: 6802445, UDI-DI: 10758750012343; Product Code: 6900440, UDI-DI: 10758750033201; Serial Numbers: J46000208 J46000019 J46000861 J46000209 J46000020 J46000910 J46000240 J46000150 J46000983 J46000497 J46000162 J46001029 J46000864 J46000247 J46001041 J46000865 J46000296 J46001042 J46000898 J46000393 J46001119 J46001030 J46000410 J46001210 J46001064 J46000478 J46001217 J46001125 J46000499 J46001289 J46001132 J46000531 J46001541 J46001252 J46000564 J46001565 J46001293 J46000617 J46001609 J46001462 J46000772 J46001629 J46001703 J46000788 J46001639 J46001704 J46000800 J46001645 J46000820 J46001680 J46000859

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide. Global Distribution.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Dec 20, 2023
Recall initiation date
Oct 10, 2023
Center classification date
Dec 13, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records