VITROS 3600 Immunodiagnostic System Product Code 6802783 and Refurbished Product Code 6802914 Running Software Versio... — FDA record Z-0543-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Oct 10, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
VITROS 3600 Immunodiagnostic System Product Code 6802783 and Refurbished Product Code 6802914 Running Software Version 3.8.0
Reason
During planned monitoring of data post-release of Software Version 3.8.0 an increased occurrence of software issues relating to unresponsive subsystems, unresponsive user interface (sometimes referred to as a screen freeze), and/or printer disconnections was identified. The consequence of these issues would be a delay in reporting results due to the required system shutdowns to recover the system. This delay in reporting results could occur during any assay, including which the speed of the result is deemed critical or STAT and could result in a delay of treatment in a patient. Due to the release strategy of this software version, the user could not update the system but a field engineer had to manually install each affected system.
Classification
Class II
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Quantity
63 units
Agency-reported status
Ongoing
Source record identifier
Z-0543-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityProduct Code: 6802783, UDI-DI: 10758750002979; Product Code: 6802914, UDI-DI: 10758750007103; Serial Numbers: J36000109 J36000294 J36001012 J36001489 J36000292 J36000340 J36001024 J36001495 J36000484 J36000348 J36001027 J36001513 J36000489 J36000446 J36001051 J36001515 J36000955 J36000511 J36001088 J36001519 J36001151 J36000517 J36001108 J36001528 J36001167 J36000580 J36001111 J36001535 J36001222 J36000688 J36001112 J36001570 J36001283 J36000753 J36001117 J36001573 J36001330 J36000861 J36001237 J36001575 J36001348 J36000877 J36001245 J36001579 J36000920 J36001246 J36001581 J36000923 J36001401 J36001582 J36000929 J36001403 J36001583 J36000933 J36001427 J36001600 J36000941 J36001451 J36001602 J36000942 J36001488 J36001618 J36001000

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide. Global Distribution.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Dec 20, 2023
Recall initiation date
Oct 10, 2023
Center classification date
Dec 13, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.