VITROS Chemistry Products CRP Slides- In vitro diagnostic quantitatively measures C-reactive protein (CRP) concentrat... — FDA record Z-0535-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Oct 11, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
VITROS Chemistry Products CRP Slides- In vitro diagnostic quantitatively measures C-reactive protein (CRP) concentration in serum and plasma Model/Catalog Number: 1926740
Reason
May Generate Negatively Biased Results When Testing Samples with High Concentrations of CRP, showing less than upper AMI of 90 mg/L (in the 70-90 mg/L range) while the samples having higher than 90 mg/L. User no way to know if the sample has to be diluted to measure again.
Classification
Class II
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Quantity
7717 units
Agency-reported status
Ongoing
Source record identifier
Z-0535-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI/GTIN 10758750000296 Coating 1221, expiration date 10/01/2024 Coating 1222, expiration date 01/01/2025

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution in the states of AZ, FL, ID, IL, KS, MN, MT, NY, PA, UT and the countries of Australia, Belgium, Bermuda, Brazil, Canada L3R 0B8, Chile, China, Colombia, Denmark, France, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore 609917, Spain, Sweden, The Netherlands, United Kingdom.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Dec 4, 2024
Recall initiation date
Oct 11, 2024
Center classification date
Nov 25, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.