Recall archive / FDA device enforcement
(1) Medline Pacer Implant Pack, REF DYNJ59194A, 4 kits/case, sterile; and (2) Medline Thoracotomy Morristown, REF DYN... — FDA record Z-0533-2025
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Oct 15, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- (1) Medline Pacer Implant Pack, REF DYNJ59194A, 4 kits/case, sterile; and (2) Medline Thoracotomy Morristown, REF DYNJ904220F, 1 kit/case, non-sterile.
- Reason
- Some of the suture packages within the convenience kits contain the incorrect needle type/size and suture diameter/length.
- Classification
- Class II
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Quantity
- 360 kits
- Agency-reported status
- Ongoing
- Source record identifier
- Z-0533-2025
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | (1) REF DYNJ59194A - Lot numbers 24DLA916, exp. 12/31/2025; and 24BLA030, exp. 11/30/2025; UDI-DI each 10193489573480, UDI-DI case 40193489573481; and (2) REF DYNJ904220F - Lot number 24DLA070, exp. 11/30/2024, UDI-DI each 10195327362300, UDI-DI case 40195327362301. |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
US Nationwide distribution in the states of AL, GA, MA, NJ, and PA.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Dec 4, 2024
- Recall initiation date
- Oct 15, 2024
- Center classification date
- Nov 25, 2024
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.