PIVO Blood Collection Device 20G, REF: 202-0005 — FDA record Z-0527-2022

Reported terminated

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Nov 10, 2021. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
PIVO Blood Collection Device 20G, REF: 202-0005
Reason
Incomplete sealing of blood collection device unit packages may produce a breach in the sterile barrier enclosing the product. This could lead to microbial contamination of the device prior to use. The health effects range from no clinical effect up to bloodstream infection.
Classification
Class II
Recalling firm
Velano Vascular
Quantity
123,700
Agency-reported status
Terminated
Source record identifier
Z-0527-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI: 00850984007027. Lots: 070621-02, 062221-01, 061021-02, 052721-03, 051421-02

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the states of CT, CO, UT, PA, IL, WA, CA, NJ, MI, FL, NC.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Feb 2, 2022
Recall initiation date
Nov 10, 2021
Center classification date
Jan 25, 2022
Termination date
Jul 31, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records