ARTIS icono (ceiling configuration), Fluoroscopic x-ray system; Model no. 11328100 — FDA record Z-0523-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Oct 17, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
ARTIS icono (ceiling configuration), Fluoroscopic x-ray system; Model no. 11328100
Reason
Potential hardware issue for ARTIS icono ceiling system: collision sensor may falsely signal non-existing collisions for up to half an hour when the system is switched off for a longer time (e.g., overnight) and switched on. During this time, system movement is only possible with the "override" function with very slow speed and without further collision protection. This may lead to temporary unavailability of the system for diagnostic and therapeutic usage.
Classification
Class II
Recalling firm
Siemens Medical Solutions USA, Inc
Quantity
38 worldwide, 11 US
Agency-reported status
Ongoing
Source record identifier
Z-0523-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI 04056869295923 Serial Numbers: All SN's associated with material number 11328100

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide - Worldwide Distribution: Domestic distribution to FL GA IA IL IN MI MN MO MT PA TX. Foreign distribution to Australia Canada Switzerland China Germany Denmark Egypt Spain Finland United Kingdom Israel Netherlands Sweden South Africa

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Dec 20, 2023
Recall initiation date
Oct 17, 2023
Center classification date
Dec 9, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.