Brand Name: MST Duet Reusable Handle for Single Use Heads Product Name: Reusable Handle for Single Use Heads (Forceps... — FDA record Z-0521-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Sep 17, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Brand Name: MST Duet Reusable Handle for Single Use Heads Product Name: Reusable Handle for Single Use Heads (Forceps and Scissors) Model/Catalog Number: DFH-1028 Software Version: N/A Product Description: A MST Single Use Forceps or Scissors System consists of a reusable handle and one of a number of single use interchangeable heads. MST Single Use Forceps and Scissor heads are shipped ready to use sterile in a tray with Tyvek lid within a sterile pouch. MST Duet Reusable Handle for Single Use Forceps & Scissor Heads are supplied in a non-sterile condition and must be sterilized prior to and after every use. The blue sleeve the handle is packaged in is not intended to be sterilized and does not maintain sterility. Component: Yes. The MST Duet¿ Reusable Handle for Single Use Heads is the reusable handle component which connect to either Single Use Forceps or Scissors heads.
Reason
Due to pin coming off of handles.
Classification
Class II
Recalling firm
MicroSurgical Technology Inc
Quantity
24
Agency-reported status
Ongoing
Source record identifier
Z-0521-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityLot Code: Product Codes: HNF or HNR (Regulation Number 886.4350). Either code can apply as the Duet Handle Assembly can be used with MST Duet Ophthalmic Scissors or MST Duet Forceps. UDI: 00844685005536 Lot #: 160458. Lot number is a sequential number generated through the internal SAP. All finished goods can be effectively traced by the lot number as either the finished good part or the sub part used. Expected life: The number of handle reprocessing cycles is not limited as long as the post-cleaning and pre-use inspections determine that the product is still functional and shows no signs of damage or corrosion.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

U.S.: CT, GA, MN, MT and TX. O.U.S.: Netherlands 24 total devices = 6 devices distributed in the United States and 18 devices distributed in Germany.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Dec 4, 2024
Recall initiation date
Sep 17, 2024
Center classification date
Nov 22, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records