Recall archive / FDA device enforcement
MEDLINE convenience kits labeled as: 1) BREAST BIOPSY PACK-LF, REF DYNJ0160598C; 2) BIOPSY BREAST PACK, REF DYNJ65361... — FDA record Z-0515-2026
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Sep 30, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- MEDLINE convenience kits labeled as: 1) BREAST BIOPSY PACK-LF, REF DYNJ0160598C; 2) BIOPSY BREAST PACK, REF DYNJ65361A; 3) BREAST BIOPSY/MASTECTOMY, REF DYNJ908839A.
- Reason
- Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.
- Classification
- Class II
- Recalling firm
- Medline Industries, LP
- Quantity
- 301 units
- Agency-reported status
- Ongoing
- Source record identifier
- Z-0515-2026
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | 1) REF DYNJ0160598C: UDI/DI 10195327304270 (each), 40195327304271 (case), Lot Numbers: 23DDC250, 23GDA351; 2) REF DYNJ65361A: UDI/DI 10195327415051 (each), 40195327415052 (case), Lot Numbers: 24BMC296; 3) REF DYNJ908839A: UDI/DI 10195327431822 (each), 40195327431823 (case), Lot Numbers: 23GBL904, 23GBU981, 23HBF736, 23HBO368, 24BBP268, 24CBO698, 24EBS782. |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
US Nationwide distribution.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Nov 26, 2025
- Recall initiation date
- Sep 30, 2025
- Center classification date
- Nov 14, 2025