Artis pheno with software VE20C- diagnostic imaging angiography system Model: 10849000 — FDA record Z-0515-2022

Reported terminated

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Oct 12, 2021. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Artis pheno with software VE20C- diagnostic imaging angiography system Model: 10849000
Reason
Four potential software issues with Artis pheno and Artis icono systems with software VE20C: 1. Updated calibration data not saved with measurement after scene+/-; 2. No x-ray possible, system shutdown/restart might be required during intervention; 3. Corrupted Image during Roadmap; 4. Unintended shutdown of Imaging System with UPS (Uninterruptable Power Supply) option
Classification
Class II
Recalling firm
Siemens Medical Solutions USA, Inc
Quantity
12 units
Agency-reported status
Terminated
Source record identifier
Z-0515-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilitySerial Numbers: 164719 164712 164722 164723 164728 164711 164718 164703 164721 164726 164724 164725 UDI: 04056869046877

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Feb 2, 2022
Recall initiation date
Oct 12, 2021
Center classification date
Jan 21, 2022
Termination date
Apr 11, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records