Voyager Rollator-intended as a mechanical walker Model Number: RLEU10xx (RLEU10BL, RLEU10PK, RLEU10WT) RLEU10BL Voyag... — FDA record Z-0510-2022

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jan 10, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Voyager Rollator-intended as a mechanical walker Model Number: RLEU10xx (RLEU10BL, RLEU10PK, RLEU10WT) RLEU10BL Voyager Rollator Cobalt Blue UPC: 815067078066; RLEU10PK Voyager Rollator Rose Gold UPC: 815067078073; RLEU10WT Voyager Rollator Ice Palace UPC: 815067078080
Reason
Wheel spoke may crack causing the wheel to separate from the axle may result in patient injuries, including bruising and lacerations requiring stitches.
Classification
Class II
Recalling firm
Compass Health Brands (Corporate Office)
Quantity
3734 units
Agency-reported status
Ongoing
Source record identifier
Z-0510-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityLot/Serial # RM2005000001-RM21004005000 (May 2020 through April 2021 manufacture dates)

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Nationwide

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jan 26, 2022
Recall initiation date
Jan 10, 2022
Center classification date
Jan 19, 2022

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.