MC3 VitalFlow Console, REF 58100; Blood pump of ecmo — FDA record Z-0509-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Oct 3, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
MC3 VitalFlow Console, REF 58100; Blood pump of ecmo
Reason
As of August 6, 2025, Medtronic has received eleven reports of VitalFlow Consoles displaying an E70 error code during normal operation. When this occurs, the touch screen may become temporarily unresponsive and go blank for up to two minutes before recovering to full functionality. Importantly, the console continues to maintain set pump speed and function throughout.
Classification
Class II
Recalling firm
Medtronic Perfusion Systems
Quantity
167 units
Agency-reported status
Ongoing
Source record identifier
Z-0509-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI/DI 10850007732247, Serial Numbers: 000023, 000047, 000073, 000100, 000126, 000150, 000173, 000197, 000223, 000024, 000048, 000075, 000101, 000127, 000151, 000174, 000202, 000224, 000025, 000051, 000076, 000102, 000128, 000152, 000177, 000203, 000225, 000027, 000052, 000078, 000103, 000129, 000153, 000179, 000204, 000228, 000028, 000054, 000079, 000104, 000130, 000154, 000180, 000205, 000229, 000029, 000055, 000080, 000106, 000131, 000155, 000181, 000206, 000277, 000030, 000056, 000081, 000108, 000132, 000156, 000182, 000207, 000284, 000031, 000058, 000082, 000109, 000133, 000157, 000183, 000208, 000032, 000059, 000083, 000110, 000134, 000159, 000184, 000209, 000033, 000060, 000084, 000112, 000135, 000161, 000185, 000210, 000034, 000061, 000085, 000115, 000136, 000162, 000186, 000211, 000035, 000063, 000087, 000116, 000140, 000163, 000187, 000212, 000037, 000065, 000088, 000117, 000141, 000165, 000188, 000213, 000038, 000066, 000089, 000118, 000142, 000166, 000189, 000214, 000039, 000067, 000090, 000119, 000143, 000167, 000190, 000215, 000040, 000068, 000092, 000120, 000145, 000168, 000191, 000216, 000043, 000069, 000095, 000121, 000146, 000169, 000192, 000217, 000044, 000070, 000096, 000122, 000147, 000170, 000193, 000218, 000045, 000071, 000097, 000124, 000148, 000171, 000194, 000219, 000046, 000072, 000099, 000125, 000149, 000172, 000195, 000221.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Nov 26, 2025
Recall initiation date
Oct 3, 2025
Center classification date
Nov 14, 2025