What the record says

Product
Panorama 1.0T HFO, Magnetic Resonance (MR) System; Model Numbers 781250 and 781350. Philips Magnetic Resonance (MR) systems are Medical Electrical Systems for use as a diagnostic device.
Reason
Firm was informed of an event in which the structural integrity of the MR system components failed due to unintended excessive pressure buildup of helium gas during a magnet quench. If the system is unable to contain the pressured helium gas due to an unforeseen blockage in the system, risks to patients and/or operators that may lead to injury or death include chemical exposure (i.e. helium gas) that may expand to surrounding rooms, contusion caused by debris, and asphyxia.
Classification
Class I
Recalling firm
Philips North America Llc
Quantity
150 US; 192 ROW
Agency-reported status
Ongoing
Source record identifier
Z-0505-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityAll serial numbers

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Domestic distribution to AL AZ CA CO DE FL GA IA IL IN KS KY LA MA MD MI MO NE NJ NV NY OH OK OR PA SC TN TX UT VA VT WA WI WV International distribution to Argentina Austria Canada China Colombia Croatia Czech Republic Denmark Ecuador Egypt France Germany Greece Italy Kuwait Libya Netherlands NO93 Panama Poland Portugal Qatar Russian Fed. Saint Kitts Saudi Arabia Spain Sweden Switzerland Syria Thailand T¿rkiye United Kingdom Utd.Arab.Emir. Venezuela

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Dec 20, 2023
Recall initiation date
Nov 17, 2023
Center classification date
Dec 14, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.