VITROS Immunodiagnostic Products High Sample Diluent A Reagent Pack — FDA record Z-0503-2022

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Nov 22, 2021. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
VITROS Immunodiagnostic Products High Sample Diluent A Reagent Pack
Reason
Lower than expected results for VITROS¿ Immunodiagnostic Products TSH Reagent Pack when diluted with VITROS¿ Immunodiagnostic Products High Sample Diluent A Reagent Pack Lot 2190.
Classification
Class II
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Quantity
1416 units
Agency-reported status
Ongoing
Source record identifier
Z-0503-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityLot # 2190; Product code(unique Device Identifier): 8430373 (10758750004898)

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, DC, DE, FL, GA, IA, IL, IN, KY, LA, MD, ME, MI, MO, MS, MT, NC, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WI, WV and the countries of Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, Singapore, United Kingdom, France, Germany, Italy, Spain, Portugal, Russia, Denmark, Norway, Sweden, Belgium, The Netherlands.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jan 26, 2022
Recall initiation date
Nov 22, 2021
Center classification date
Jan 18, 2022

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.