OrganOx metra -Intended as a Normothermic Machine Perfusion system of donor transport livers. Model/Catalog Number: D... — FDA record Z-0501-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Oct 15, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
OrganOx metra -Intended as a Normothermic Machine Perfusion system of donor transport livers. Model/Catalog Number: D0003 The OrganOx metra is a transportable device intended to be used to sustain donor livers destined for transplantation in a functioning state for a total preservation time of up to 12 hours.
Reason
Battery movement led to damage of one of the battery cells causing it to short to the battery retention bracket
Classification
Class II
Recalling firm
OrganOx Ltd
Quantity
20 units
Agency-reported status
Ongoing
Source record identifier
Z-0501-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 05060462240029 Serial numbers 511, 512, 514, 515, 516, 517, 518, 519, 520, 522, 523, 524, 525, 527, 528, 529, 530, 531, 533, 535, 536, 537

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

California, Texas, Ohio, Florida, New York, Massachusetts, Virginia, Oklahoma, DC, South Carolina and Georgia.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Nov 27, 2024
Recall initiation date
Oct 15, 2024
Center classification date
Nov 21, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.