Draeger Fabius MRI Anesthesia Machine, Catalog Number 8607300 — FDA record Z-0501-2022

Reported terminated

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Dec 16, 2021. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Draeger Fabius MRI Anesthesia Machine, Catalog Number 8607300
Reason
26 Fabius MRI Anesthesia machines were distributed with internal hoses routing Oxygen to the Oxygen Flush and Oxygen Flow Control interchanged. If the hospital O2 Central supply pressure is not stable, the O2 fresh gas flow could deviate from the adjusted value. If the deviation is not noticed, insufficient oxygenation of the patient may occur.
Classification
Class II
Recalling firm
Draeger Medical, Inc.
Quantity
5
Agency-reported status
Terminated
Source record identifier
Z-0501-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI 04048675041498 Serial Numbers ASPJ-0019 ASPJ-0020 ASPJ-0021 ASPK-0001 ASPK-0002

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the states of NY, PA, and WI.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jan 26, 2022
Recall initiation date
Dec 16, 2021
Center classification date
Jan 14, 2022
Termination date
Apr 17, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.