Recall archive / FDA device enforcement
Draeger Fabius MRI Anesthesia Machine, Catalog Number 8607300 — FDA record Z-0501-2022
Reported terminated
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Dec 16, 2021. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Draeger Fabius MRI Anesthesia Machine, Catalog Number 8607300
- Reason
- 26 Fabius MRI Anesthesia machines were distributed with internal hoses routing Oxygen to the Oxygen Flush and Oxygen Flow Control interchanged. If the hospital O2 Central supply pressure is not stable, the O2 fresh gas flow could deviate from the adjusted value. If the deviation is not noticed, insufficient oxygenation of the patient may occur.
- Classification
- Class II
- Recalling firm
- Draeger Medical, Inc.
- Quantity
- 5
- Agency-reported status
- Terminated
- Source record identifier
- Z-0501-2022
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | UDI 04048675041498 Serial Numbers ASPJ-0019 ASPJ-0020 ASPJ-0021 ASPK-0001 ASPK-0002 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
US Nationwide distribution in the states of NY, PA, and WI.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Jan 26, 2022
- Recall initiation date
- Dec 16, 2021
- Center classification date
- Jan 14, 2022
- Termination date
- Apr 17, 2023