Siemens Atellica CH Reaction Ring Cuvette Segment-has cuvettes that contain and incubate the sample reaction volumes ... — FDA record Z-0496-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Oct 21, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Siemens Atellica CH Reaction Ring Cuvette Segment-has cuvettes that contain and incubate the sample reaction volumes and present them to each processing step. Siemens Material Number(SMN): 11099326 All lots starting with N15 or 15 are expected to have the defect.
Reason
A manufacturing defect causes Atellica CH Reaction Ring Cuvette Segments with lot numbers N15 on the packaging or 15 on the underside of the cuvette segment have a potential for imprecision and erroneously depressed or elevated QC and patient results for seven (7) analytes: Ammonia, Acetaminophen, Alanine Aminotransferase (ALT), Hemoglobin A1c, Creatinine Kinase (CK), Cystatin C, and Salicylate.
Classification
Class II
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
9729 units
Agency-reported status
Ongoing
Source record identifier
Z-0496-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityLot # UDI N1527024 (01)00630414596143(10)N1527024(17)20291002; N1526424 (01)00630414596143(10)N1526424(17)20291001; N1525924 (01)00630414596143(10)N1525924(17)20290924 ; N1525324 (01)00630414596143(10)N1525324(17)20290918 ; N1524224 (01)00630414596143(10)N1524224(17)20290911; N1523724 (01)00630414596143(10)N1523724(17)20290904; N1523024 (01)00630414596143(10)N1523024(17)20290827; N1521824 (01)00630414596143(10)N1521824(17)20290814; N1521424 (01)00630414596143(10)N1521424(17)20290808; N1519324 (01)00630414596143(10)N1519324(17)20290719; N1518924 (01)00630414596143(10)N1518924(17)20290716; N1518224 (01)00630414596143(10)N1518224(17)20290710; N1518324 (01)00630414596143(10)N1518324(17)20290710; N1516224 (01)00630414596143(10)N1516224(17)20290621; N1515624 (01)00630414596143(10)N1515624(17)20290612; N1514924 (01)00630414596143(10)N1514924(17)20290605; N1514124 (01)00630414596143(10)N1514124(17)20290530; N1513324 (01)00630414596143(10)N1513324(17)20290522; N1512824 (01)00630414596143(10)N1512824(17)20290515; N1512424 (01)00630414596143(10)N1512424(17)20290510; N1511224 (01)00630414596143(10)N1511224(17)20290506; N1510724 (01)00630414596143(10)N1510724(17)20290423; N1510124 (01)00630414596143(10)N1510124(17)20290419.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide and Worldwide distribution: Foreign: Algeria Algeria Argentina Australia Austria Bahrain Bangladesh Belgium Brazil Bulgaria Burkina Faso Canada Chile China Colombia Congo, Republic C¿te dIvoire Croatia Cyprus Czech Republic Denmark Ecuador Egypt Estonia Finland France Germany Greece Hong Kong Hungary India Iran, Islamic Republic of Iraq Ireland Israel Italy Japan Jordan Kenya Kuwait Latvia Libya Malaysia Mexico Moldova, Rep. of Morocco Nepal Netherlands Norway Oman Pakistan Paraguay Peru Philippines Poland Portugal Qatar Republic Korea Romania Russian Federation Saudi Arabia Serbia Singapore Slovakia Slovenia South Africa Spain Sweden Switzerland Taiwan, Province of China Thailand Tunisia Turkey Uganda Ukraine United Arab Emirates United Kingdom Great Britain Uruguay Vatikancity Vietnam

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Nov 27, 2024
Recall initiation date
Oct 21, 2024
Center classification date
Nov 20, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.