GLOBAL UNITE STD STEM SZ 8 Part Number: 110008100 — FDA record Z-0484-2022

Reported terminated

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Dec 7, 2021. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
GLOBAL UNITE STD STEM SZ 8 Part Number: 110008100
Reason
Stems may have an undersized spigot bore diameter therefore separating the proximal body from the stem may not be possible without surgical delay and/or bone damage
Classification
Class II
Recalling firm
DePuy Orthopaedics, Inc.
Quantity
432 units
Agency-reported status
Terminated
Source record identifier
Z-0484-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityLot # s 9793035, 9802134, 9802135, 9802136, 9802137, 9802138, 9802139, 9802140, 9802141, 9802142, 9802143, 9802144, 9802145, 9802146, 9802147, 9802148, 9802149, 9802150, 9802151, 9814515, 9817165, 9817166, 9817167, 9817168, 9817633, 9818331, 9824332, 9824334, 9824335, 9824336, 9824337, 9824338, 9824339, 9824340, 9824341, 9824342, 9824343, 9824344, 9824345, 9824346, 9824347, 9832804, 9851705, 9851706, 9851707, 9851708, 9862057, 9864147, 9874672, 9874674, 9874675, 9874676, 9874678, 9874682, 9874683, 9874686, 9874688, 9887446 GTIN: 10603295004042

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide and the countries of Austria, Australia, Belgium, Canada, Denmark, Estonia, Finland, France, Germany, Greece, Ireland, Israel, Italy, Netherlands, New Zealand, Norway, Poland, South Africa, Sweden, Switzerland, Thailand, United Kingdom.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jan 19, 2022
Recall initiation date
Dec 7, 2021
Center classification date
Jan 11, 2022
Termination date
May 17, 2024