Brand Name: Olympus Single Use Single Use Electrosurgical Snare SD-400 Product Name: Single Use Single Use Electrosur... — FDA record Z-0483-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Oct 6, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Brand Name: Olympus Single Use Single Use Electrosurgical Snare SD-400 Product Name: Single Use Single Use Electrosurgical Snare Model/Catalog Number: SD-U400U-15 Product Description: These instruments have been designed to be used with an Olympus endoscope for the removal and/or cauterization of diminutive polyps, sessile polyps, pedunculated polyps, and tissue from within the GI tract.
Reason
Sterility compromised from small holes that may occur during packaging of the device pouches into the shelf carton and/or during transportation of the shelf carton.
Classification
Class II
Recalling firm
Olympus Corporation of the Americas
Quantity
18174 units
Agency-reported status
Ongoing
Source record identifier
Z-0483-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityModel: SD-400U-15; Lot Numbers: 27V, 28V, 29V, 2XV, 2YV, 2ZV, 31V, 32V, 33V, 34V, 35V, 36V, 37V, 38V, 39V, 3XV, 3YV, 3ZV, 41V, 42V, 43V, 44V, 45V, 46V, 47V, 48V, 49V, 4XV, 4YV, 4ZV, 51V, 52V, 53V, 54V, 55V, 56V UDI: 04953170408250

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide and the countries of GU, and MP.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Nov 19, 2025
Recall initiation date
Oct 6, 2025
Center classification date
Nov 7, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.