Foot Controlled Electrosurgical Suction Coagulator, Electrosurgical Suction Coagulator 10FR, Catalog Number 130187 — FDA record Z-0474-2023

Reported terminated

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Nov 18, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Foot Controlled Electrosurgical Suction Coagulator, Electrosurgical Suction Coagulator 10FR, Catalog Number 130187
Reason
Suction ports may potentially be occluded on the affected devices. If this occurs, the affected devices will not be able to perform the suction function during use, which could cause a delay in procedure and could require another device to be used.
Classification
Class II
Recalling firm
ConMed Corporation
Quantity
23,700 US; 5,225 OUS
Agency-reported status
Terminated
Source record identifier
Z-0474-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 30653405059930 Lot codes 202204294 202205134 202205164 202206084 202206154 202206204 202206304 202207074 202207084 202208014 202208044 202208084

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Domestic distribution nationwide. Foreign distribution to Colombia, Ecuador, Nicaragua, Panama, Thailand, Australia, Canada, and Lebanon.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Dec 21, 2022
Recall initiation date
Nov 18, 2022
Center classification date
Dec 9, 2022
Termination date
Apr 13, 2026

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.